createaccountlogin

Dissolved Gas Content Sensors - CARBOTEC

 carbotec tr

Operating Principle
Measuring Range:                         0 - 10 g/litre CO2
Accuracy:                                         +/- 0.05 g/litre CO
2 
Reproducibility:                              +/- 0.01g/litre CO2

Temperature Compensation:     PT100
Temperature Range:                     -10 - +100°C

Pressure:                                          max. 10 bar g.
Enclosure Rating:                          IP65

Description

            The Carbotec TR-PT in-line transmitter measures continuously and exactly the content of dissolved gas in liquids, for example for monitoring the carbon dioxide content in carbonated drinks.

 

            For determining the CO2-content of a carbonated liquid, every 20 seconds a 25ml sample is isolated from the main stream. The pressure of this sample is abruptly reduced to below atmospheric pressure. The resultant pressure will differ from that of a non-carbonated liquid according to the volume of dissolved CO2 in the sampled liquid. This difference in resulting pressure is the basis for calculating the CO2- content of the liquid. The temperature dependence of the measurement signal is electronically compensated by PT100. After the measurement the sample is returned to the beverage line. The CO2-concentration can be displayed in different units like e.g. g/l.

 

            The menu-driven software includes the product-specific calibration and product selection. Analogue and digital in- and outputs onboard can be used for process controlling and automation. The sensor has a hygienic construction and is cleanable with all in food- and beverage industry typical CIP-media